The Desktop Conductivity Meter EZL-DCM121 has a water-resistant membrane keypad and a large dual display. The conductivity measuring range of the model is 0.00 µS/cm to 200.0 mS/cm. At an ambient temperature of 5 °C to 35 °C, a 0 to 100 g/L TDS with a resistivity of 0 to 100 MW.cm and automatic temperature compensation ranging from 0 °C to 100 °C having an accuracy of 5 °C to 60 °C : ± 0.5 °C and others: ± 1.0 °C controls temperature variations and provides greater precision.
| Measuring Range |
0.00 to 20.00 µS/cm 20.0 to 200.0 µS/cm 200 to 2000 µS/cm 2.00 to 20.00 mS/cm 20.0 to 200.0 mS/cm |
| Resistivity | 0 to 100 MΩ.cm |
| TDS | 0 to 100 g/L |
| Salinity | 0 to 100 ppt |
| Resolution | 0.01/0.1/1 µS/cm 0.01/0.1 mS/cm |
| Accuracy |
Electrode: ± 1.0 % FS Instrument: ± 1.50 % FS |
| Temperature Compensation | 0 °C to 100 °C (Automatic) |
| Electrode Constant | 0.1/1/10 cm-1 |
| Temperature Measuring Range | 10 °C to 110 °C |
| Temperature Resolution | 0.1 °C |
| Temperature Accuracy | 5 °C to 60°C: ± 0.5 °C Others: ± 1.0 °C |
| Ambient Temperature | 5 °C to 35 °C |
| Ambient Humidity | ≤ 80 % |
| Power | DC 12V/1A |
| Dimensions (W×D×H) | 240×170×70 mm |
| Weight | 600 g |
It has a wide range of applications, including high precision conductivity measurements in scientific research, quality control, biotechnology, and fine chemistry, such as measuring the conductivity of common water solutions, high purity water, reverse osmosis water, and distilled water.