labtron.online

Water Purification System LWPS-A10

Water Purification System LWPS-A10 is a Type I water treatment system and the feed water is RO or distilled water. It mainly consists of pretreatment system, RO system, ultrapure water system conveying unit and control system. It consists of 254 nm UV lamp to remove the bacteria in the system and 185nm UV lamp to reduce TOC. Fully automatic control over the whole process, e.g. automatic water preparation, when water tank is full, automatically stop the system and cut off the power and water. Built-in RO membrane timing automatic flush procedures, and manual RO membrane cleaning button is available.

Water quality One stage (Type I)
Feed water RO water /Distilled water
Water temperature: 5 ~ 40℃
Resistivity 18.25 MΩ.cm @ 25 ℃
TOC < 5ppb
Bacteria < 0.1 cfu/ml
Endotoxin < 0.001 EU/ml
Particle ( > 0.22μm) <1/ml
Absorbance (254nm,1cm optical distance) < 0.001 EU/ml
Reactive Silica(SiO2) < 0.01 ppm
RNases < 0.01 ng/ml
DNases < 4 pg/μl
Temperature 5 ℃ ~ 40 ℃
Humidity 10 % ~ 80 %
Water production rate Type 1 water flow rate: 1-1.5 L/min
Micro filtration Millipore Millipak 0.22μm membrane or hollowed fiber membrane filter
Power supply 230 V/50 Hz 150 W
Main body Dimension (L × W × H) 340 × 550 × 530 mm (bench type)

  • Millipore BioPak UF membrane to reduce the pyrogen, RNase, DNase

  • Ultra-purification system: Millipore ion exchange resin cartridge and special flow design, high ultra-purify efficiency

  • Effectively remove the trace ions in water; guarantee the stability of water quality

  • Micro filtration: Making the particles content (larger than 0.22μm) is <1/ml and bacteria less than 0.1CFU/ml

  • Built-in manual water cycling program and ultrapure water auto-cycle program to keep the water quality of ultrapure water

  • Water resistance on-line monitoring, realizing man-machine conversation

  • System can be installed on the lab bench

  • 30L PE water reservoir with vent valve

Used for Medical device injection to human body, DNA/RHA preparation, DNA testing, HPLC (test and analysis of amino acid, penicillin; Organic ingredients), GCMS, ICP, IC, gas chromatography, AAS etc.

This water purification system provides Type II level water with consistency and reliability. It removes approximately 98% of inorganic ions and 99 % of all dissolved organic substances

UV water purification systems destroy the DNA of micro-organisms making the water contamination free. Water produced from UV water purification is contamination free water which is potable for consumption

Copyright © 2026 Labtron. All rights reserved.